Search result
Farah,
K. S.; Sneddon, J. Optimization of
a simultaneous multi-element
atomic-absorption spectrometer.
Talanta 40(6):879-882 (1993).
- Abstract:
- A variable-size simplex
procedure was used to
optimize the operating
conditions for the
simultaneous determination of
Cu, Zn, Mn and Fe by flame
AAS. Seven factors were
optimized including the slit
width, the air-to-fuel ratio,
the height above the burner
head and the four
hollow-cathode lamp currents.
Results confirm that
multi-element atomic
absorption requires a
compromise in operating